Specialized Tools for Electrophysiology and Cell Biology Research

O2/CO2 Controllers

Product Summary

Choose between fully manual or fully digital mixers.

Manual Three Gas Mixers

Three Gas Mixer / (3GF Mixer-Hypoxia) Manual CO2 - Manual N2 - Manual Air

  • CO2 range 0 - 15%
  • O2 range 2 - 20%
  • Set point resolution 1%
  • PTFE membrane filter (0.2 mm pores

The 3GF Mixer-Hypoxia is a 3 gas mixer (Carbon Dioxide, Nitrogen, Air) which employs floating ball flow meters.


Three Gas Mixer / (3GF Mixer-Hyperoxia) Manual CO2 - Manual N2 - Manual Air

  • CO2 range 0 - 15%
  • O2 range 2 - 20%
  • Set point resolution 1%
  • PTFE membrane filter (0.2 mm pores

The 3GF Mixer-Hypoxia is a 3 gas mixer (Carbon Dioxide, Nitrogen, Air) which employs floating ball flow meters.

O2/CO2 Controllers

Bold Line Three Gas Mixer Multi-user

CO2-O2 UNIT-BL CP

Designed to supply desired Air-CO2 or N2-O2-CO2 mixture at controlled pressure.

  • Controls O2, N2, and CO2
  • Ideal for hypoxia and hyperoxia experiments.
  • Gas flow rates digitally controlled.
  • Operation is via an easy to use touch screen interface, OKO-Touch.
  • An optional Smart Box data logger and web server allows operation via any web-enabled device.
O2/CO2 Controllers

The Bold Line Three Gas Mixer is designed to replace pre-mixed tanks or compressed gas lines and a single unit can supply up to 10 microscope incubators.

The device can mix two or three gases, according to operator's need.

A long life zirconium oxide sensor precisely measures the O2 concentration, while a non dispersive InfraRed (NDIR) dual wave length detector sets the CO2 concentration. Operation is via an easy to use touch screen interface, OKOTouch.

An optional Smart Box data logger and web server allows operation via any web-enabled device. The system may also be operated via a PC using OKO-Control 2.0 software or Nis Elements software.

Sensor calibration is made with an external meter or calibration gas.

Note: Order Single floating ball flowmeters along with the Multi-user controller, one for each incubator to be connected.

OKOLab StageTop Incubator System Overview

The Oko Touch is necessary for operation of the controller.

OKOLab StageTop Incubator System Overview

Bold Line CO2 - O2 Controllers

The Bold Line combination CO2 - O2 gas controllers are ideal for hypoxia and hyperoxia experiments.

  • We are the Authorized Distributor for Okolab products in the US
  • Controls O2, N2, and CO2
  • Ideal for hypoxia and hyperoxia experiments.
  • Gas flow rates digitally controlled.
  • Operation is via an easy to use touch screen interface, OKO-Touch.
  • An optional Smart Box data logger and web server allows operation via any web-enabled device.

CO2-O2 UNIT-BL [0-10;1-18] and [0-20;1-95]

The Bold Line combination CO2 - O2 gas controllers are ideal for hypoxia and hyperoxia experiments. Two versions are available, CO2; 0-10% and O2; 1-18%, and CO2; 0-20% and O2; 1-95%.

Gas flow rates are digitally controlled. A long life zirconium oxide sensor precisely measures the O2 concentration, a non dispersive InfraRed (NDIR) dual wave length detector sets the CO2 concentration.

OKOLab StageTop Incubator System Overview

Operation is via an easy to use touch screen interface, OKO-Touch. An optional Smart Box data logger and web server allows operation via any web-enabled device. The system may also be operated via a PC using OKO-Control 2.0 software or Nis Elements software.

Sensor calibration is made with an external meter or calibration gas.

Required: Oko Touch

(necessary for operation of controller)

OKOLab StageTop Incubator System Overview

Gas Controllers

Bold Line Specifications
CO2-O2 UNIT-BL [0-20;1-95]

Output pressure 1-4 atm (absolute pressure)
Pressurized Vessel 5 L
Operating mode Adds CO2 and air to N2
Concentration range CO2 range: 4-8%, O2 range: 3-7%
CO2 Accuracy ± 5% of CO2 concentration, i.e. ± 0.25% at 5%CO2
O2Accuracy 0.1% at 5%O2
Set Point Resolution 0.1%
CO2 Repeatability Better than 0.1%
O2 Repeatability 0.05% of O2
Total flow rate 0.5 Nl/min (at 4 ata)
CO2 consumption (at 6% of CO2): 0.042 Nl/min
Air consumption (at 6% of CO2 and 5% O2): 0.167 Nl/min
N2 consumption (at 5% of 02) 0.5 Nl/min
CO2 sensor 10 year-life Non Dispersive InfraRed (NDIR) dual wave length detector
O2 sensor 10 year-life zirconium oxide sensor
Filtering device PTFE membrane with 0.2 μm pores

 



Bold Line CO2 - O2 Controller Specifications
CO2-O2 UNIT-BL [0-10;1-18]

 

Operating mode Adds CO2 and N2 to Air
Concentration range CO2 range: 0-10%, O2 range: 1-18%
CO2 Accuracy ± 5% of CO2 concentration, i.e. ± 0.25% at 5%CO2
O2Accuracy 0.1% at 5%O2
Set Point Resolution 0.1%
CO2 Repeatability Better than 0.1%
O2 Repeatability 0.05% of O2
Total flow rate 0.4 Nl/min - constant
Outlet pressure Ambient
CO2 consumption (at 5%) 0.02 Nl/min
N2 consumption (at 5% of 02) 0.3 Nl/min
CO2 sensor 10 year-life Non Dispersive InfraRed (NDIR) dual wave length detector
O2 sensor 10 year-life zirconium oxide sensor
Filtering device PTFE membrane with 0.2 μm pores

 


CO2-O2 UNIT-BL [0-20;1-95]

 

Operating mode Mixes CO2, O2 and N2
Concentration range CO2 range: 0-20%, O2 range: 1-95%
CO2 Accuracy ± 0.3 % at 5% CO2 and 5% O2
O2Accuracy ± 0.2% at 5% CO2 and 5% CO2
Set Point Resolution 0.1%
Total flow rate 0.35 Nl/min - constant
Outlet pressure Ambient
CO2 consumption (at 5% of CO2): 0.0175 Nl/min
O2 consumption (at 5% of O2): 0.0175 Nl/min
N2 consumption (at 5% of CO2 and at 5% of O2): 0.315 Nl/min
Digital Flow Meters CMOS Sensors
CO2 sensor 10 year-life Non Dispersive InfraRed (NDIR) dual wave length detector
O2 sensor 10 year-life zirconium oxide sensor
Filtering device PTFE membrane with 0.2 μm pores

Order No. Model No. Product Description USA Price Order
Manual Mixer
64-1997 3GF-Mixer-Hypoxia 3 Gas mixer, Air Nitrogen and CO2, O2-2-20% Login Login
64-1998 3GF-Mixer-Hyperoxia 3 Gas mixer, Air Nitrogen and CO2, O2-40-85% Login Login
Bold Line Multi-User
64-2004 CO2-O2 Unit-BL CP Bold Line N2 / CO2 / O2 gas controller, multi-user. CO2 range from 4-8%, O2 range from 3-7% Login Login
64-2965 Oko-Touch Touch screen display for Bold Line controllers Login Login
Bold Line CO2 - O2
64-2003 CO2-O2 Unit-BL[1-20:1-95] Bold Line CO2 / O2 controller for hypoxia and hyperoxia. CO2 to 20%, O2 to 95% Login Login
Accessories and Replacement Parts
64-1999 OKO-AP Air Pump Login Login
64-2008 SM-BL Smart Box (data logger, web server, and remote assistant) for Bold Line controllers Login Login
64-2962 OKO-AP-BL Bold Line Air Pump, max pressure 300 mbar Login Login

Need support?

Want to know more about our quality products or need assistance in technical support?

TALK TO OUR EXPERTS

NEWSLETTER

For the most up-to-date information on Warner Instruments news and technologies, sign-up for our newsletter.