Specialized Tools for Electrophysiology and Cell Biology Research

(TC-120) Thermocirculator

Product Summary

  • Electronic controls
  • Internal or external circulation modes
  • Facility for simple recalibration by the user
  • Suitable for both flammable and non-flammable liquids
  • Display can be switched between °C and °F
  • Minimum stray magnetic field
  • Automatic self-tests
  • Overtemp cutouts and alarms

TC120 Heated Circulating Bath

  • Operating temperature range: ambient +15 to 99°C
  • Powerful 16L/min pump allows temperature-controlled fluid to be pumped to external devices
  • Simple-to-use rotary control allows quick access to user interface
  • Adjustable over-temperature cutout
  • Simple 2-point recalibration
  • Low liquid protection float switch
  • Three user-adjustable pre-set temperatures
  • Visual and audible alarms alert you when your attention is required
  • Clear 4-digit display - easy to read from a distance

Versatile mid-range model with digital thermostatic control unit and stainless steel tank and a comprehensive specification to suit most applications for high precision temperature control.

Grant TC120 series systems combine a digital heating circulator with an economy plastic tank.

The Grant TC120 heating circulator incorporates Intelligent Control Optimisation"(ICO) for adaptive intelligent PID temperature control and includes a powerful integral pump, making TC120 systems suitable both for immersing samples and circulating temperature controlled fluid to external devices. It includes a convenient timer function for reaction timing and a variable high temperature alarm setting. The temperature range and heater power are automatically limited according to the liquid type selected.

Note: Accessory cooling is required for operation at or below ambient temperature Note: Alternatively, the Grant Optima TC120 heating circulator can also be used to regulate the temperature in vessels other than Grant tanks. Such vessels must have vertical sides and a maximum wall thickness of 35 mm for rectangular tanks, 30 mm for circular tanks (300 mm diameter) and a capacity of up to 50 litres. The control unit is attached to other vessels by means of a clamp.

Applications

Heated circulating baths provide a source of precision heating and cooling for many routine and sensitive analytical procedures. They are suitable for use as both open and closed loop circulators (i.e. remote vessel open or closed). Alternatively, the Grant FH series of flow heaters (closed circulators) can be used. For more powerful heating requirements, e.g. i.e. above +200ºC, see Grant high temperature baths or contact Grant for advice.

TC120 Series Heated Circulating Baths

Temperature Range Ambient +15 to 99°C
Working area (l x w) 120 x 150 mm
Min/max liquid depths 80/140 mm
Inner tank dimensions (l x w x h 240 x 160 x 150 mm
Overall dimensions incl. controller (l x w x h) 390 x 200 x 380 mm
Stability (DIN 12876) @ 70°C ± 0.05
Uniformity (DIN 12876) @ 70°C ± 0.1
Setting resolution 0.1
Display 4 digit LED
Timer function 1 to 9999 mins
No. preset temperatures 3
Two point re-calibration yes
Safety adjustable cut-out
Fluid level - float switch yes
Heater power 120 V kW 1.4
Heater power 230 V kW 1.3
Electrical power 120 V kW 1.5(50-60 Hz)
Electrical power 230 V kW 1.4 (50 Hz)
Height above tank rim 200 mm
Depth below tank rim 135 mm

Customers from Europe and Australia can use the Multi Channel Systems - Online Shop to order Warner products or create quotes.

Order No. Model No. Product Description USA Price Order
72-4924 TC-120 Thermocirculator, 5 Liter Plastic Bath with Lid, 120 V, 60 Hz Login Login
72-4921 TC-120 Thermocirculator, 5 Liter Plastic Bath with Lid, 220 V, 50 Hz Login Login
64-2244 WB-KIT TC-120 Thermocirculator Tubing Kit Login Login

Need support?

Want to know more about our quality products or need assistance in technical support?